A - 15
Note: The specifications above are based on a new battery, and at 20ºC±5 °C of ambient temperature.
Battery LI34I001A charge
time
Charged by the equipment
connected to the AC power
Less than 2 hours to 90% and less than 3 hours to 100%
with equipment power off;
Less than 6 hours to 90% and less than 9 hours to 100%
with equipment power on.
Charged by the charger station Less than 3.5 hours to 90% and less than 4 hours to
100%.
Battery LI24I004A run time
Operating
mode
One new fully
charged
battery
Two new fully
charged batteries
Testing condition
Monitoring ≥6 h ≥12 h The equipment is configured
with a 5-lead ECG, Resp, SpO
2
,
2-channel temp and CO
2
, NIBP
measurements set at an
interval of 15 minutes.
WI-Fi is
disabled. The screen brightness
is set to the factory default
without recording.
Defibrillation ≥200
discharges
≥400 discharges 360J discharge, the equipment
is configured with a 5-lead ECG,
Resp, SpO
2
, 2-channel temp,
NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to
the factory default without
recording.
≥300
discharges
≥600 discharges 200J discharge, the equipment
is configured with a 5-lead ECG,
Resp, SpO
2
, 2-channel temp,
NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to
the factory default without
recording.
Pacing ≥4.5 h ≥9 h 50  load impedance, pacing
rate 80ppm, pacing output
60mA. The equipment is
configured with a 5-lead ECG,
Resp, SpO
2
, 2-channel temp,
NIBP and CO
2
. WI-Fi is disabled.
The screen brightness is set to
the factory default without
recording.
Battery LI34I001A run time
Operating
mode
One new fully
charged
battery
Two new fully
charged batteries
Testing condition
Monitoring ≥5 h ≥10 h With a typical configuration,
the equipment operates with
ECG only. WI-Fi is disabled. The
screen brightness is set to the
factory default without
recording.
Defibrillation ≥100
discharges
≥200 discharges 360J discharge at intervals of
one minute, without recording
Pacing ≥3 h ≥6 h 50  load impedance, pacing
rate: 80ppm, pacing output
60mA, without recording
Battery fuel gauge 5 LEDs indicating the current battery charge level
Shutdown delay At least 20 minutes of monitoring and six 360J discharges (after the low battery alarm
occurs)