3-29
The message [Calibration Completed!] is displayed after a successful calibration. If the
calibration failed, the prompt [Calibration Failed!] will be displayed.
C.O. Test
Tools required:
Medsim300B Patient simulator, or equivalent equipment
C.O. adapter box (for 300B)
PiCCO test cable (PN: 040-001301-00)
Follow this procedure to perform the test:
1. Connect the patient simulator and the C.O. module using a C.O. trunk cable and a C.O.
adapter box.
2. Set the blood temperature (BT) to 37ºC on the patient simulator and check the
temperature value displayed on the monitor is 37 ± 0.1ºC.
3. In the [CCO Setup] menu, select [PiCCO Guide>>], and check if the value for
[Cat.Type] is [PV2015L20].
4. Turn the injectate temperature (TI) knob on the C.O. adapter box to set the TI to 20 ±
1ºC for the patient simulator and C.O. to 5L/min.
5. In the PiCCO screen, select [Start] to start C.O. measurement. As soon as the prompt
[Inject XXml] is displayed, adjust TI to 4 ± 1ºC, and then quickly back to 20 ± 1ºC.
Simutaneously press the button on the simulator that corresponds to 5L/min. The whole
procedure shall be finished in 10 s.
6. Verify that the C.O. value displayed on the monitor is correct.

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals