EasyManua.ls Logo

Mindray BeneView T8 - Page 75

Mindray BeneView T8
234 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-41
External Device Setup Remark
Draeger Evita 2 dura/ Evita
4/ Evita XL
Protocol: Medibus
Baud Rate: 19200
Parity: even
Stop Bits: 1
Interval: ---( Evita 2 dura)
/
Newport E360 Protocol: Newport
The following information is for
further reference:
Baud Rate: 38400 bps
Word Length: 8 bits
Parity: NONE
Stop Bits: 1
Puritan Bennett 840
Baud Rate: 38400
Word Length: 8 bits
Parity: NONE
/
Hamilton G5(protocol
Polling)
Protocol: Polling.
The following information is for
further reference:
Baud Rate: 9600 bps
Word Length: 7 bits
Parity: even
Stop Bits: 2
Hamilton C2(protocol
Polling)
Protocol: Polling.
The following information is for
further reference:
Baud Rate: 9600 bps
Word Length: 7 bits
Parity: even
Stop Bits: 2
Hamilton Galileo(protocol
Polling)
Not required.
The following information is for
further reference:
Baud Rate: 9600 bps
Word Length: 7 bits
Parity: even
Stop Bits: 2
Hamilton G5(protocol
Block)
Protocol: Block.
The following information is for
further reference:
Baud Rate: 38400 bps
Word Length: 8 bits
Parity: none
Stop Bits: 1

Table of Contents