3-42
External Device Setup Remark
Ohmeda Avance/Aisys Not required.
The following information is for
further reference:
Baud Rate: 19200 bps
Word Length: 7 bits
Parity: odd
Stop Bits: 1
Ohmeda Aestiva 7100
/7900
Not required.
The following information is for
further reference:
Baud Rate: 19200 bps
Word Length: 7 bits
Parity: odd
Stop Bits: 1
Drager Fabius GS/Plus/Tiro
Protocol: Medibus
Baud Rate: 9600
Word Length: 7 bits
Parity: even
Stop Bits: 1
/
Drager Primus
Protocol: Medibus
Baud Rate::9600
Word Length: 8 bits
Parity: even
Stop Bits: 1
/
TCM CombiM/TCM
TOSCA
Protocol: Monlink.
The following information is for
further reference:
Baud Rate: 9600 bps
Word Length: 8 bits
Parity: even
Stop Bits: 1
TOF-Watch SX Not required.
The following information is for
further reference:
Baud Rate: 19200 bps
Word Length: 8 bits
Parity: none
Stop Bits: 1