33 - 2 BeneVision N Series Patient Monitor Operatorās Manual
33.3.1 Hemodynamic Parameters
33.3.1.1 Hemodynamic Parameters from the ICG module
The following table lists hemodynamic parameters from the ICG module.
33.3.1.2 Hemodynamic Parameters from PiCCO module and ScvO
2
module
The following table lists the hemodynamic parameters from the PiCCO module and ScvO
2
module.
Abbreviation Full Spelling Unit
Output C.O. Cardiac Output L/min
C.I. Cardiac Index L/min/m
2
SV Stroke Volume ml
SVI Stroke Volume Index ml/m
2
HR Heart Rate bpm
Contractility LCW Left Cardiac Work kgĀ·m
LCWI Left Cardiac Work Index kgĀ·m/m
2
LVSW Left Ventricular Stroke Work gĀ·m
LVSWI Left Ventricular Stroke Work Index gĀ·m/m
2
ACI Acceleration Index /100s
2
PEP Pre-ejection Period ms
VI Velocity Index /1000s
STR Systolic Time Ratio No unit
LVET Left Ventricular Ejection Time ms
Preload Volume TFI Thoracic Fluid Index ī
TFC Thoracic Fluid Content /kī
CVP Central Venous Pressure mmHg
PAWP Pulmonary Artery Wedge Pressure mmHg
Afterload SVR Systemic Vascular Resistance DS/cm
5
SVRI Systemic Vascular Resistance Index DSĀ·m
2
/cm
5
PVR Pulmonary Vascular Resistance DS/cm
5
PVRI Pulmonary Vascular Resistance Index DSĀ·m
2
/cm
5
Art-M Mean Arterial Pressure mmHg
Art-S Systolic Arterial Pressure mmHg
Art-D Diastolic Arterial Pressure mmHg
Abbreviation Full Spelling Unit
Output CCO Continuous Cardiac Output L/min
CCI Continuous Cardiac Index L/min/m
2
SV Stroke Volume ml
SVI Stroke Volume Index ml/m
2
HR Heart Rate bpm