BeneVision N Series Patient Monitor Operator’s Manual 39 - 3
39.3.4 The Discharge Tab
39.3.5 The Location Tab
Data Transfer Strategy Always Ask The monitor needs a data transfer strategy if the 
monitor detects that the patient demographic 
information in the monitor and the N1/T1/MPM is 
not consistent. 
•
Always Ask: always prompts a dialog box to 
ask for strategy.
•
Continue Patient in Module: continues to 
use the patient information in the N1/T1/
MPM. The monitor discharges the patient, and 
automatically admits a new patient and 
copies all the patient information from the 
N1/T1/MPM.
•
Continue Patient in Monitor: continues to 
use the patient information in the monitor. 
The monitor deletes all the patient 
information in the N1/T1/MPM and copies the 
parameter settings in the monitor to the N1/
T1/MPM.
Tabular Trends Selected Selects what kind of data will be transferred.
Event
Event Detail
Waveform
12-Lead
Menu Item Default Setting Description
Menu Item Default Setting Function
Auto Discharge When Power Off Never Automatically discharges the patient when the 
monitor is turned off for the designated period of 
time. 
Never: not discharge a patient no matter for how 
long the monitor has been switched off.
Auto delete discharged patients on storage 
space is full
On /
Prompt on patient auto deleted On
On: an alarm is issued when the monitor 
automatically deletes earlier discharged patients.
Alarm on storage is nearly full Med Selects whether an alarm is issued when the 
monitor memory is very low and the priority of 
this alarm.
Clear All Patient Data / Deletes all patient information and data. 
Clearing patient data will discharge the current 
patient.
Menu Item Default Setting Description
Location 1 - Location 10 / Selects where the patient goes after patient 
monitoring stops.