BeneVision N Series Patient Monitor Operator’s Manual C - 9
C.1.6 IBP Default Settings
Initial Pressure Adult: 160 mmHg
Pediatric: 140 mmHg
Neonate: 90 mmHg
Interval OR: 5 min
NICU: 30 min
Other departments: 15 min
Start Mode Clock
NIBP End Tone Off
Venipuncture Pressure Auto
Display Format Sys/Dia (Mean)
Alarm Limit Display On
Display PR Off
Item Default Setting
Item Default Setting
IBP-S Alarm switch On
High limit
â–  Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
A
dult: 160 mmHg
P
ediatric: 120 mmHg
Neonat
e: 90 mmHg
â–  PA
A
dult: 35 mmHg
Pediatric and neonate: 60 mmHg
Low limit
â–  Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
A
dult: 90 mmHg
P
ediatric: 70 mmHg
Neonat
e: 55 mmHg
â–  PA
A
dult: 10 mmHg
P
ediatric and neonate: 24 mmHg
Priority Med
Alarm Outputs Off