A - 32 BeneVision N Series Patient Monitor Operator’s Manual
A.15.15 RM Specifications
Technique Diff-Pressure flow measurement technology
Flow
Measurement range Adult/pediatric:
Neonate:
± (2 to 120) L/min
± (0.5 to 30) L/min
Accuracy Adult/pediatric:
Neonate:
1.2 L/min or ±10% of the reading,
whichever is greater
0.5 L/min or ±10% of the reading,
whichever is greater
Resolution 0.1 L/min
Paw
Measurement range -20 to 120 cmH
2
O
Accuracy ±3%×reading
Resolution 0.1 cmH
2
O
MVe/MVi
Measurement range Adult/Pediatric:
Infant:
2 to 60 L/min
0.5 to 15 L/min
Resolution 0.01 L/min if airflow is less than 10 L/min, 0.1 L/min if airflow is equal to or
more than 10 L/min
Accuracy ±10% × reading
TVe/TVi
Measurement range Adult/Pediatric:
Infant:
100 to 1500 ml
20 to 500 ml
Resolution 1 ml
Accuracy Adult/pediatric:
Infant:
±10% or 15 ml, whichever is greater
±10% or 6 ml, whichever is greater
awRR
Measurement range 4 to 120 rpm
Resolution 1rpm
Accuracy 4 to 99 rpm
100 to 120 rpm
±1 rpm
±2 rpm
Calculated Parameters
Measurement range Measurement accuracy
I:E 4:1 to 1:8 Not specified.
FEV1.0% 0 to 100% Not specified.
Pmean 0 to 120 cmH
2
O±10%
PEEP 0 to 120 cmH
2
O Not specified.
PEF 2 to 120 L/min ±10%
PIF 2 to 120 L/min ±10%
PIP 0 to 120 cmH
2
O±10%

Table of Contents

Other manuals for Mindray BeneVision N12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N12 and is the answer not in the manual?

Mindray BeneVision N12 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N12
CategoryMedical Equipment
LanguageEnglish

Related product manuals