BeneVision N Series Patient Monitor Operator’s Manual A - 31
EtCO
2
High (low limit + 2) to 99 mmHg 1 mmHg
EtCO
2
Low 1 to (high limit - 2)mmHg
FiCO
2
High 0 to 99 mmHg
FiCO
2
Low 1to (high limit - 2)mmHg
EtO
2
High (low limit + 2%) to 100% 1%
EtO
2
Low 0% to (high limit - 2)%
FiO
2
High (low limit + 2%) to 100%
FiO
2
Low 18% to (high limit - 2)%
EtN
2
O High (low limit + 2) to 100% 1%
EtN
2
O Low 0 to (high limit - 2)%
FiN
2
O High (low limit + 2) to 100%
FiN
2
O Low 0 to (high limit - 2)%
EtHal/Enf/Iso High (low limit + 0.2) to 5.0% 0.1%
EtHal/Enf/Iso Low 0 to (high limit - 0.2)%
FiHal/Enf/Iso High (low limit + 0.2) to 5.0%
FiHal/Enf/Iso Low 0 to (high limit - 0.2)%
EtSev High (low limit + 0.2) to 8.0% 0.1%
EtSev Low 0 to (high limit - 0.2)%
FiSev High (low limit + 0.2) to 8.0%
FiSev Low 0 to (high limit - 0.2)%
EtDes High (low limit + 0.2) to 18.0% 0.1%
EtDes Low 0 to (high limit - 0.2)%
FiDes High (low limit + 0.2) to 18.0%
FiDes Low 0 to (high limit - 0.2)%

Table of Contents

Other manuals for Mindray BeneVision N12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N12 and is the answer not in the manual?

Mindray BeneVision N12 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N12
CategoryMedical Equipment
LanguageEnglish

Related product manuals