39 - 10 BeneVision N Series Patient Monitor Operator’s Manual
39.7 The Review Settings
39.7.1 The Tabs Tab
39.7.2 The Event Tab
39.7.3 The Arrhy Mark Tab
From the Arrhy Mark page, you can define whether the compressed ECG waveform segments for arrhythmia 
events are marked with a specific background color.
PAWP Timeout 15 min The measurements become outline fonts after a 
preset time. This avoids older values being 
misinterpreted as current measurements.
C.O. Timeout 15 min
NIBP Timeout 15 min
TempIF Timeout 30 min
CO2 Flow Rate for Neo (For Sidestream CO
2
 
Module Without O
2
)
90 ml/min Selects flow rate when using the sidestream CO
2
 
without the O
2
 monitoring function to monitor a 
neonatal patient.
Outline Font for Suspected Values On Selects whether unreliable HR, SpO
2
, and BIS 
measurements are displayed in outline font. This 
prevents unreliable measurements from being 
misinterpreted as normal measurements,
Menu Item Default Setting Description
Menu Item Default Setting Description
Tabular Trends Selected Hides the trends you do not need to review if 
deselected.
Graphic Trends
Events
Full Disclosure
OxyCRG
12-Lead ECG
ST
Menu Item Default Setting Description
Lethal Lock Selected Selects what kind of events will be locked. Locked 
events will not be deleted.
UnselectedHigh
Med
Low
Rename Event On Selects whether arrhythmia events can be 
renamed.