EasyManuals Logo

Mindray BeneVision N12 Operator's Manual

Mindray BeneVision N12
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #472 background imageLoading...
Page #472 background image
C - 8 BeneVision N Series Patient Monitor Operator’s Manual
NIBP-D Alarm switch On
High limit Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
Low limit Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority Med
Alarm Outputs Off
NIBP-M Alarm switch On
High limit Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
Low limit Adult: 60 mmHg
Pediatric: 50 mmHg
Neonate: 25 mmHg
Priority Med
Alarm Outputs Off
NIBP-S Extreme Alarm switch Off
High limit Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low limit Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 35 mmHg
Priority High
Alarm Outputs Off
NIBP-D Extreme Alarm switch Off
High limit Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low limit Adult: 35 mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority High
Alarm Outputs Off
NIBP-M Extreme Alarm switch Off
High limit Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low limit Adult: 45 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
Priority High
Alarm Outputs Off
Item Default Setting

Table of Contents

Other manuals for Mindray BeneVision N12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N12 and is the answer not in the manual?

Mindray BeneVision N12 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N12
CategoryMedical Equipment
LanguageEnglish

Related product manuals