33 - 4 BeneVision N Series Patient Monitor Operator’s Manual
33.3.1.3 Hemodynamic Parameters from CCO/SvO
2
 module
The following table lists the hemodynamic parameters from the Vigilance II, Vigileo, EV1000, and HemoSphere 
monitors. 
Abbreviation Full Spelling Unit Applicable Device
Output CCO Continuous Cardiac Output L/min All
CCI Continuous Cardiac Index L/min/m
2
All
C.O. Cardiac Output L/min Vigilance II, EV1000, 
HemoSphere
CI Cardiac Index L/min/m
2
Vigilance II, EV1000, 
HemoSphere
SV Stroke Volume ml All
SVI Stroke Volume Index ml/m
2
All
PR Pulse Rate bpm EV1000
Contractility ESV End Systolic Volume ml Vigilance II
ESVI End Systolic Volume Index ml/m
2
Vigilance II
RVEF Right Ventricular Ejection 
Fraction
% Vigilance II, 
HemoSphere
GEF Global Ejection Fraction % EV1000
CFI Cardiac Function Index l/min EV1000
Preload Volume HR Heart Rate bpm Vigilance II
CVP
C
entral Venous Pressure cmH
2
O, kPa, or mmHg All
MAP Mean Arterial Pressure mmHg or kPa Vigilance II, 
HemoSphere
EDV End Diastolic Volume ml Vigilance II
EDVI End Diastolic Volume Index ml/m
2
Vigilance II
SVV Stroke Volume Variation % Vigileo, EV1000
GEDV Global End Diastolic Volume ml EV1000
GEDI Global End Diastolic Volume 
Index
ml/m
2
EV1000
ITBV Intrathoracic Blood Volume ml EV1000
ITBI Intrathoracic Blood Volume 
Index
ml/m
2
EV1000
PPV Pulse Pressure Variation % HemoSphere
Afterload SVR Systemic Vascular Resistance DS/cm
5
 or kPa-s/l All
SVRI Systemic Vascular Resistance 
Index
DS•m
2
/cm
5
 or kPa-s-m
2
/l All
eArt-S Systolic Arterial Pressure from 
the EV1000 monitor
mmHg or kPa EV1000
eArt-M Mean Arterial Pressure from 
the EV1000 monitor
mmHg
 or kPa EV1000
eArt-D Diastolic Arterial Pressure 
from the EV1000 monitor
mmHg or kPa EV1000