7-8 7 Service and Maitenance
7.3 Weekly Maintenance
7.3.1 Cleaning Sample Probe
WARNING
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.
BIOHAZARD
In case your skin contacts the sample, control or calibrator, follow
laboratory safety procedure and consult a doctor.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
1
Place the Power to OFF.
2
Remove the cover from the sample/reagent disk.
3
Remove the sample/reagent disk.
4
Pull the probe arm to the highest point by hand. Rotate the probe arm
to move the probe to a position above the sample/reagent
compartment and convenient to operate.