2 System Performance and Workflow 
2-2
Reaction liquid volume 
BS-120/BS-130: 180-500µl 
BS-180/BS-190: 150-500µl 
Reaction temperature 
37℃ 
Test scope 
Clinical chemistries, immunoassays, TDM (Treatment Drug Monitoring) 
Auto dilution 
Dilution ratio 4~150; dilution is done in reaction cuvette. 
Operation mode 
System  and  tests  are  configured  via  the  operating  software.  Profiles  and 
calculation tests are allowed. 
Calibration method 
Linear (one-point, two-point and multi-point), Logit-Log 4p, Logit-Log 5p, Spline, 
Exponential, Polynomial and Parabola 
Programming Controls 
Westgard Multi-rule, Cumulative sum check, Twin plot 
Data processing 
Capable of storing and outputting various data and tables/graphs, and calculating 
among different tests 
Dimensions 
  l×b×h:690 mm×570 mm×595 mm. 
Weight 
≤75kg 
Emergent samples 
Emergent samples can be inserted during test at any time. 
Network connection 
Able to be connected with LIS (Laboratory Information Management System) 
2.1.2    Specifications for Sample System 
Sample tube type 
Microtube: Φ10×37mm, Φ12×37mm; 
Blood collecting tube: Φ12×68.5mm, Φ12×99, Φ12.7×75, Φ12.7×100, Φ13×75, 
Φ13×100;