8-10 8 Test and Maintenance Software
Debugging start 1: Sending instruction: <DVON>, returning instruction <ISE!>
Debugging start 2: Sending instruction: <MVON>, returning instruction <ISE!>
Debugging shutdown: Sending instruction: <DVFF>, returning instruction <ISE!>
Program check: Sending instruction: <CKSM>, returning instruction <ISV cccc>
Read electrode output voltage: Sending instruction: <RDMV>, returning instruction
<AMV Li xxx.x Na xxx.x K xxx.x Cl xxx.x>
Read reagent chip: Sending instruction: <DLRD_pp>, returning instruction <DSN
s0s0s1s1……..s6s6s7s7 c>
Write reagent chip: Sending instruction: <DLWR_pp_aa_dd>
8.2.7 Temperature Unit
Click Temp Unit on the higher-left area and then you can see the fuction buttons of the
reaction unit.
Select the Reaction checkbox, and then click Enable/Disable temperature control to
turn on the reaction disk temperature control. Disselect the Reaction checkbox, and then
click Enable/Disable temperature control to turn off the reaction temperature control.
Reagent PreHeat is to control the reagent preheating. The control method is the same
with Reaction.
Length of Sensor Line: orginal line- 3.5cm extra line- 51.5 cm. Both are fixed
parameters.
Orginal Data: Input 0, 37 and 100 to the T column, and corresponding resistance values
in the R column. Click Caculate to get R0, A, B.
When the reaction disk temperature sensor is replaced, the R0, A, B should be
recalculated by using the function provided by Orginal Data. The R0, A, B should be
configured in PARA and Speed-Temperature Unit.
When temperature collection board is replaced, AD should be reconfigured in PARA
and Speed-Temperature Unit.
The interface to configure temperature control parameters is shown in the Figure 8-10.
Reagent preheating sensor: keep the default.
Reaction disk sensor: input the calculated R0, A, B in the first three column; input AD
value in the fourth column ( found in the temperature collection board. If the board is not
replaced, leave it unmodified.)
Target temp. of reaction disk[0.01
]: keep 37.8 as default.
Reagent preheating temp. control: keep 12/10/2 as default.
Target temp. for reagent preheating: keep 45 as default.
Reaction temp. control: keep 100/250/50 as default.