Setting up the System 
CAUTION 
 
To define such parameters as sample volume, reagent volume and 
wavelength, follow the instructions in this manual and the package 
insert of the reagents. 
Samples 
CAUTION 
 
Use samples that are completely free of  insoluble substances like 
fibrin,  or  suspended  matter;  otherwise  the  probe  may  be 
blocked.Medicines, anticoagulants or  preservative in samples  may 
lead to unreliable results. 
Medicines, anticoagulants or preservative in the samples may lead to 
unreliable results. 
Hemolysis, icterus or lipemia in the samples may lead to unreliable 
test results, so a sample blank is recommended. 
Store  the  samples  properly.  Improper  storage  may  change  the 
compositions of the samples and lead to unreliable results. 
Sample volatilization may lead to unreliable results. Do not leave the 
sample open for a long period.
 
Some  samples  may  not  be  analyzed  on  the 
BS-120/BS-130/BS-180/BS-190 based on parameters the reagents 
claim  capable  of  testing.  Consult  the  reagent  manufacturer  or 
distributor for details. 
Certain samples need to be processed before being analyzed by the 
system. Consult the reagent manufacturer or distributor for details. 
The system has specific requirements on the sample volume. Refer 
to this manual for details. 
Load the sample to correct position on the sample disk before the 
analysis begins; otherwise you will not obtain correct results.