8 Test and Maintenance Software
8-23
Figure 8-20 Macro Instruction Screen
8.6.2.2 Setting up Macro Instruction Name
1) Enter the administrator interface and click Setup, input serial port password “bs120”
and then click OK;
2) Select New.
3) Enter the name of the new macro instruction in the popup dialog box, and then select
OK. The name appears in the drop-down list box above Index and Instruction.
8.6.2.3 Adding Single instruction
1) Select or create a macro instruction name.
2) Select an instruction.
Method: Select the desired unit and the instruction type in the Instruction area. All
qualified instructions of the unit are displayed in the lower table.
3) Enter the instruction values.
Enter instruction value in the Value column according to the other information in
the table.
The last two lines of the Value column can be left blank.
The Value column for the reserved instructions can be left blank.
4) Verify the instructions.