9 Troubleshooting   
9-4 
Severity: Invalidating module (Reserved) 
Level
Description  Actions taken by the system 
12  Errors  to  invalidate 
ISE tests 
The  ISE  tests  that  are  affected  by  bubbled  sample  or  go  out  of  measurement  range  due  to 
abnormal(insufficiently-aspirated) sample are invalidated and then run aga
continuously for 3 times due to the same cause, the system will skip all ISE tests related to the 
sample during the current batch of tests. 
 
Severity: Forbidding module (Reserved) 
Level
Description  Actions taken by the system 
13  Errors  to  forbid 
sample  bar  code 
scanning 
When the sample bar code reader goes wrong and cannot scan sample bar code label normally, 
the system will  not try again during the measurement. The sample positions should be set up 
manually. Only when reconnected and proved to be normal, the sample bar code reader can start 
working. 
14  Errors to forbid ISE
When errors (like calibration slope out of range) occur due to failed ISE component, ISE tests will 
be forbidden. If such errors occur during measurement, the system will invalidate the failed ISE 
tests and skip all other ISE tests in current batch. If such errors occur in idle status, the system 
will not include ISE tests once starting analysis. 
 
Severity: Errors to forbid LIS (Reserved) 
Level
Description  Actions taken by the system 
15  Errors to forbid LIS  When  the  LIS  host  goes  wrong  or  the  network  connection  and  settings  are  improper,  the 
system cannot download sample information from or send test results to LIS. You can use all 
functions associated to LIS only after reconnecting the system to LIS successfully.