9 Troubleshooting
9-8
400000590000
Operating software
error: Lamp intensity on
the low side
Light intensity is weak Install new lamp or adjust photoelectric gain.
400000600008
Operating software
error: Lamp intensity too
low. Can't test
Light intensity is weak
Check whether the lamp is installed properly and the
status of the filter wheel is alright.If yes, it might be that the
lamp reaches its service life and replace it.
400000610008
Operating software
error: Dark current
checking failed
Dark current too strong
Check the status of the lamp to avoid the system being
affected by instense light.
400000620008
Operating software
error: Dark current too
large
Dark current too strong
Check the status of the lamp to avoid the system being
affected by instense light.
400000630008
Operating software
error: Both AD values
are too similar
Turning on/off light error
Rerun the test after checking the status of the lamp and
filter wheel.
400000650008
Operating software
error: Lamp turning on
failed
Turning on/off light error
Adjust the lamp by using Alignment. If the error remains,
check the connection of the lamp.
400000660008
Operating software
error: Lamp turning off
failed
Turning on/off light error
Adjust the lamp by using Aligment. If the error remains,
check the connection of the lamp.
400000690008
Operating software
error: Reaction
temperature too high
Reaction temperature too
high or temperature control
error
Shut down Power. Check whether the reaction disk
temperature sensor cable and temperature sensor are ok.
Check whether the heater and overheat-protection switch
are ok.
400000700008
Operating software
error: Reaction
temperature too low
Reaction disk temperature
control error
Check the status of the reaction disk temperature sensor
and whether the heater is on.The reaction disk
temperature wil return to normal, when the sensor is ok
and the heater is turned on for a certain period.
400000710000
Operating software
error: Temperature
fluctuation
Temperature fluctuation
Check the reaction disk cuvette replacement cover to
make sure it is closed.
400000720000
Operating software
error: Sending buffer
overflows
Sending buffer overflows Please contact the R&D engineer.