9 Troubleshooting 9-17
100680210005
Sample unit result error:
Sample probe horizontal
movement error. Cannot
reach the home position
Sample probe horizontal
movement error. Cannot
reach the home position
Check the connection between probe horizontal movement
motor and the sensor. Check whether the sample probe
horizontal movement is blocked. Execute mechanical
resetting.
100680210015
Sample unit result error:
Sample probe horizontal
movement error. Cannot
move away from the
home position
Sample probe horizontal
movement error. Cannot
leave the home position
Check the connection between probe horizontal movement
motor and the sensor. Check whether the sample probe
horizontal movement is blocked. Execute mechanical
resetting.
100680210055
Sample unit result error:
Sample probe horizontal
movement error. Step
missing
Sample probe horizontal
movement error. Step
missing.
Check the movement of the sample probe by using
Alignment. Check whether the horizontal movement of the
sample probe is blocked. Execute mechanical resetting.
100680210065
Sample unit result error:
Sample probe horizontal
movement error.
Disabled
Sample probe horizontal
movement error.
Forbidden.
Execute mechanical resetting.
100680220005
Sample unit result error:
Sample probe vertical
movement error. Cannot
reach the home position
Sample probe vertical
movement error. Cannot
reach the home position
Check the connection between probe vertical movement
motor and the sensor. Check whether the sample probe
vertical movement is blocked. Execute mechanical
resetting.
100680220015
Sample unit result error:
Sample probe vertical
movement error. Cannot
move away from the
home position
Sample probe vertical
movement error. Cannot
leave the home position
Check the connection between probe vertical movement
motor and the sensor. Check whether the sample probe
vertical movement is blocked. Execute mechanical
resetting.
100680220025
Sample unit result error:
Sample probe vertical
movement error. Trying
to move away from the
initial limit position
Sample probe vertical
movement error. Trying to
move away from the initial
limit position
Execute mechanical resetting. If the error remains, check
the status of the sensor.