9 Troubleshooting   
9-20 
100650030005 
Reaction  unit  result 
error:  Mechanical 
resetting error 
Mechanical resetting error
Check the motor/sensor/cable. Run mechanical resetting. 
If  the  error  remains,  restart  the  analyzing  unit  and  the 
operating software. Rerun mechanical resetting again.   
100650040005 
Reaction  unit  result 
error:  Invalid  status. 
Self-check 
Self-checking;  Other 
operations not supported 
Execute  the  previous  operation  after  waiting  for  30-60 
seconds. If there is no response for a long time, execute 
mechanical resetting or restart the analyzing unit and the 
operating software. 
100650040015 
Reaction  unit  result 
error:  Invalid  status. 
Error 
Reaction unit error; current 
operations not supported 
Execute failure recovery. Check the motor/sensor/cable. If 
the error remains, please contact the R&D engineer.   
100650040025 
Reaction  unit  result 
error:  Invalid  status. 
Waiting for handshake 
Unit busy. No response 
Execute  the  previous  operation  after  waiting  for  30-60 
seconds. If there is no response for a long time, execute 
mechanical resetting or restart the analyzing unit. Rerun 
the operating software. 
100650050005 
Reaction  unit  result 
error:  Unit  busy.  No 
response 
Unit busy. No response 
Execute  the  previous  operation  after  waiting  for  30-60 
seconds. If there is no response for a long time, execute 
mechanical resetting or restart the analyzing unit and the 
operating software. 
100650090005 
Reaction  unit  result 
error:  Wrong 
configuration parameter 
Parameter  configuration 
error 
Configure the right parameter. 
100650110005 
Reaction  unit  result 
error: Undefined search 
Undefined search  Input the right searching parameter. 
100650130005 
Reaction  unit  result 
error:  Wrong  system 
operation parameter 
Wrong  system  operation 
parameter 
Input the right operation parameter. 
100650140005 
Reaction  unit  result 
error:  Configure 
undefined parameter 
Reaction  unit  error.  Can 
not find home position 
Check  the motor/sensor/cable. Check  the  movement  of 
the reaction disk by using Aligment. If the error remains, 
contact the R&D engineer.