9 Troubleshooting
9-30
10070001CAF5
ISE unit result error:
Calibration cycle error.
No flow
No flow; wrong electrode
installation
Replace the calibrator A and B, and then recalibrate;
check the installation of the electrodes.
10070001CAM5
ISE unit result error:
Calibration cycle error
Electrodes damaged
Check the status of the electrodes and replace the
damaged one.
10070001CAQ5
ISE unit result error:
Calibration cycle error.
Calibration value saving
failed
Software read/write error
Execute the command for 3 times. If the error remains,
contact the R&D engineer.
10070001CLA5
ISE unit result error:
Clean cycle error. Air in
calibrant A
Calibrator A bottle empty;
tube not connected;
pump A not working;
tube blocked, broken or
twisted;
fibrin and salt in the tube;
air bubble detector error;
waste pump error
Wrong electrode installation. Check the pressing plate
spring and the sealing to ensure all the electrodes and
the O-ring are in the right place.
Wash the module by using <CLEN > program. Remove
the module and then wash it. Reinstall the sensor.
Replace air bubble detection sensor;
replace the waste pump;
replace the calibrator A and recalibrate;
reconnect or replace the tube;
check the connection of the motor;
replace the pump box; replace the motor;
replace the tube.
10070001CLC5
ISE unit result error:
Clean cycle error. Air in
cleaner
No ISE wash solution
air bubble detector error;
waste pump error
Replace air bubble detection sensor;
replace waste pump;
refill ISE wash solution.
10070001CLF5
ISE unit result error:
Clean cycle error. No
flow
No ISE wash solution Refill ISE wash solution