4 Units Description
4-43
4.8.3.3 Reagent Module
Figure 4-48 Reagent Module
The reagent module consists of a reagent pack and a reagent pack seat. The reagent
pack seat is fixed on the base board by the three M4X10 socket screws.
Dismounting steps:
1. Remove the ISE unit door on the left side of the analyzer.
2. Pull out the tube of the reagent module.
3. Take out the reagent pack horizontally.