Image Optimization 5-47
2. Click [SmartFace] to enter the function and the system adjust fetal face angle (fetal head
facing up and face is at the front with [Direct] to be up/down) automatically and remove
the shading obstacle data.
Parameter adjusting
Parameters under Smart Face are similar to those under Static 3D mode.
FaceContact
Description The higher the value, the more adjacent the VR image is to the fetal
face, and the more obstacles cleared.
The lower the value, the further the VR image is to the fetal face, and
the less obstacles cleared.
Operation
Rotate [FaceContact] to adjust the parameter.
The adjusting range is: -15-15 in increment of 1.
Quick Rot.
Description
Rotate the image quickly.
Operation
Adjust in VR Orientation: 0°, 90°, 180°, 270°.