Part I Service Scenarios
Maintenance 44
No. Item
Recommended
Frequency
Maintenance
Personnel
Method
Terminal
User
Engineer
5 Cleaning the probe
(head)
Each time after
being used
● ●
6 Cleaning the probe cable
and connector shell
1 time/month ● ●
7
Cleaning the holders
(including probe holder
and medical ultrasonic
couplant holder)
1 time/month ● ●
8 Cleaning the shell 1 time/month ● ●
9 Cleaning the peripherals 1 time/month ● ●
10 Checking the probe
surface
1 time/day ● ● See Part I Service
Scenarios: 2.2
Check.
11 Checking the power
cable and power plug
1 time/month ● ●
12 Checking the batteries 1 time per 3–6
months
● ●
13 Checking peripherals and
accessories
1 time/year ● ●
14 Mechanical safety check 1 time/year ● ●
15 Electrical safety check 1 time/year ●