Why does Mindray Passport 12 Medical Equipment report 'communication stopped'?
- SSarah ThomasSep 13, 2025
If your Mindray Medical Equipment reports "XX communication stopped" or some parameters cannot be used, it could be due to a defective cable inside the module, a defective parameter module, or a defective converter board inside the module. Check the cables connecting the converter board and the corresponding parameter module. If the cables are fine, consider replacing the corresponding module or the corresponding converter board.