1.Hold and pull up the rotary handle for 90 degrees.
Sodalime Canister
2.Turn the rotary handle
for 90 degrees counterclockwise.
3.
Pull out the sodalime canister from
the slot.
WARNING
Sodalime is a caustic substance and is a strong
irritant to eyes, skin and respiratory system.
Affected parts should be ushed with water. If
irritation continues after ushed by water, seek
medical assistance immediately.
Before installing the sodalime canister, inspect
the canister mouth, canister support and seal for
sodalime particles. If there is, clear it to prevent
breathing system leakage.
Clean the sodalime canister and change the
sodalime canister sponge regularly. Otherwise,
the sodalime powder built up inside the canister
will go into the breathing system.
The sodalime canister sponge must be in place
to prevent dust and particles from entering the
breathing system.
Do not reuse the sodalime canister sponge,
which must be replaced every time the sodalime
canister is replaced.
Cleaning and Disinfection
Canister seal
Canister support
Canister mouth