EasyManua.ls Logo

Mindray WATO EX-65 - Turn on the System; Input Fresh Gas

Mindray WATO EX-65
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
Turn on the System
Connect the power cord to the AC power source.Make sure that the AC power
LED is illuminated.
Set the system switch to ON.Make sure that both the operating state LED and
battery LED are illuminated (the battery is being charged or fully charged).
The alarm lamp ashes yellow and red once in turn and then a beep is given.
The display shows the start-up screen and then enters the standby screen
after half a minute.
1.
2.
3.
4.
WARNING
Do not use the anesthesia machine if it generates alarms when turned on, or
can not operate normally. Contact your service personnel or us.
Input Fresh Gas
Connect the gas supplies correctly and ensure adequate gas pressure.
You can control the O2, N2O and AIR ows in the fresh gas through the O2,
N2O and AIR ow controls. Readings of the gas ows can be seen on the
respective owmeters.
If inspiratory anesthetic agent is used, ensure that the vaporizer is correctly
mounted onto the anesthesia machine and the vaporizer conforms to the
the anesthetic agent to be used. Fill the vaporizer with the anesthetic agent
properly.
Push and turn the concentration control on the vaporizer to set the
appropriate concentration of anesthetic agent.
1.
2.
3.
4.
NOTE
:
This anesthesia machine can be mounted with vaporizers corresponding with
halothane, en urane, iso urane, sevo urane and des urane. Only one of the
ve vaporizers can be opened at a time because the vaporizers are featured
with interlock.

Other manuals for Mindray WATO EX-65

Related product manuals