22
Maintenance Schedule
:
Minimum
frequency
Maintenance
Daily
Clean the external surfaces.
21% O2 calibration (O2 sensor in breathing system).
Biweekly Drain the vaporizers and discard the agent.
Monthly
100% O2 calibration (breathing system O2 sensor).
Clear water built up inside the waterstraps of CO2 module
and AG module.
During cleaning
and setup
Inspect the parts and O-rings for damage. Replace or repair
as necessary.
Annually
Replace the seal on the vaporizer manifold and that on the
breathing system port. Contact us for details.
CO2 module calibration.
AG module calibration.
Every
three years
Replace the built-in lithium-ion batteries. Contact us for
details.
As necessary
Before installing the cylinder, use a new cylinder gasket on
cylinder yoke.
Empty the water collection cup If there is water built up in it.
Replace the sodalime in the sodalime canister if sodalime
color changes.
Replace the breathing system O2 sensor if it is damaged
(Under typical use the sensor meets speci cations for one
year.).
Replace the ow sensor if it is damaged.
Calibrate the ow sensor when any of the following occurs:
1) the ow sensor is changed; 2) the tidal volume set is not
consistent with that displayed when no leakage occurs;
3) there is a great difference between the tidal volume
measured and that displayed by the bellows scale when
fresh gas is very small.
Maintenance
NOTE
These schedules are the minimum frequency based on typical usage of 2000
hours per year. You should service the equipment more frequently if you use it
more than the typical yearly usage.

Other manuals for Mindray WATO EX-65