20
3.Ensure that the breathing system assembly has been removed correctly. 
Disinfect the breathing system assembly.
 
 Breathing system
1.
Remove the breathing system assembly 
according to the aforementioned 
procedures.Hold the breathing system 
with one hand. Pull up the locking 
catches on the breathing system 
adapter with the other hand to unlock it.   
2.Remove the breathing system from 
the breathing system adapter with both 
hands.
 
Cleaning and Disinfection
WARNING
:
Disassemble and reassemble the breathing system as described in this guide. 
For further disassembly and reassembly, contact us. Improper disassembling 
and reassembling may cause breathing system leak and compromise normal 
system use.
After the breathing system assembly is cleaned and disinfected, reinstall the 
assembly when it is fully dry.