19
Cleaning and Disinfection
1.Turn the bellows housing
counterclockwise.
2.Remove the housing.
3.Remove the folding bag from the
bellows base.
Bellows Assembly
NOTE
Before installing the bellows housing,
check that the sealing component on the
breathing system is in position. If not,
you must install the sealing component
properly before installing the bellows
housing.
When installing the bellows housing,
align the bellows housing bayonet
tabs with the slots on the breathing
system and then lower the bellows
housing. Make sure that the housing
is depressing the seal evenly. Hold
the bellows housing tightly and turn it
clockwise until it stops. Make sure that
the side of the housing marked with
scale is facing the operator.
Incorrect installation causes the folding
as shown in the gure, which leads to the
breathing system leakage. Reinstall the
bellows housing.

Other manuals for Mindray WATO EX-65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray WATO EX-65 and is the answer not in the manual?

Mindray WATO EX-65 Specifications

General IconGeneral
TypeAnesthesia Machine
ModelWATO EX-65
BrandMindray
Tidal Volume Range20 to 1500 mL
O2 Concentration21% to 100%
I:E Ratio4:1 to 1:8
Ventilation ModesVCV, PCV, SIMV, Manual, Spontaneous
Flow MeterO2, N2O
VaporizerUp to 2 vaporizers
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
MonitoringAirway Pressure, Flow, Volume
O2 Flush35 to 75 L/min
Respiratory Rate4 to 100 breaths per minute
Power Supply100 to 240 VAC, 50/60 Hz
Fresh Gas Flow0 to 15 L/min

Related product manuals