24
Failure description Possible cause Recommended action
After the calibration 
is started, the 
message of 
calibration failure 
is displayed before  
ventilation sound is 
heard.
1. The pipeline supply 
pressure is too low.
2. The bag/mechanical 
ventilation switch is not 
set to the mechanical 
position.
3. The supply pressure 
switch malfunctions.
4. The bag/mechanical 
ventilation switch 
malfunctions.
1. Check the supply pressure and 
ensure that it is not lower than 0.2 
MPa. 
2. Check the bag/mechanical 
ventilation switch and ensure 
that it is set to the mechanical 
position.
3. In case that the supply 
pressure is normal but the alarm 
[Drive Gas Pressure Low] is still 
on, contact us.
4. Switch over the bag/
mechanical ventilation switch but 
the ventilation status displayed 
in the screen upper left corner 
keeps unchanged. In this case, 
contact us.
After the calibration 
is started, the 
message of 
calibration failure 
is displayed shortly 
after ventilation 
sound is heard.
1. The fl ow sensor in 
the breathing system is 
not installed properly.
2. The check valve 
assembly in the 
breathing system is not 
installed properly.
3. The fl ow sensor 
malfunctions.
1. Check the fl ow sensor and the 
related sealing components for 
improper installation.
2. Check the check valve 
assembly and the related 
sealing components for improper 
installation.
3. Replace the fl ow sensor and 
calibrate again.
4. If the problem persists 
after taking all the mentioned 
measures, contact us.
After the calibration 
is started, the 
ventilation  sound 
is heard. After the 
calibration lasts 
about 3 minutes, 
the system prompts 
calibration failure.
During calibration, 
the gas supply or the 
breathing tubes are 
interfered so that the 
calibration data are not 
correct.
1. Remove the problem that the 
tubes are pressed. Calibrate  the 
fl ow sensor again by following 
calibration procedures.
2. If the problem persists after  
multiple calibrations, contcact us.
Flow Sensor Calibration
In case of fl ow sensor calibration failure, refer to the following table for 
troubleshooting.