EasyManua.ls Logo

Mindray Zonare ZS3

Mindray Zonare ZS3
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ZS3 Service Manual Page 145 of 295
Figure 14.10-ZS3
3. Disassemble the eight(8) screws and the UI, Thread the Gel warmer cable
through the punch hole of UI interface base and connect 4-pin connector side into
the rightmost port of the power distribution strip
Figure 14.11-ZS3
Figure 14.12-ZS3
4. Wind a circle and Tie the tail of the Gel Warmer cable and printer USB cable. (If
printer with no options)
Pan head screw M4X12
Gel warmer cable
Disassemble the eight (8) screws
and the UI
1
Gel warmer cable connect port

Table of Contents