Transducer Maintenance
Regularly check transducers for signs of wear or damage.
WARNING: Bent, broken, or missing pins on the transducer connector can
cause poor image quality, including possible mirror image artifact. Be sure to
check pins before connecting a transducer to the ZS3 system. If pins are bent,
broken, or missing, do not use the transducer and call Mindray/Zonare
Technical Support.
Figure 16.1-ZS3
Check
General Check