TABLE 2
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
Discharge(ESD)
IEC 61000-4-2
±8 kV contact;
±15 kV air
±8 kV contact;
±15kV air
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
Transient / burst
IEC 61000-4-4
±2 kV for power
supply lines;
±1 kV for
input/output lines
±2 kV for power
supply lines;
±1 kV for input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
IEC 61000-4-5
±1 kV line(s) to
line(s);
±2 kV line(s) to earth
±1 kV line(s) to line(s);
±2 kV line(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
interruptions and
voltage variation
on power supply
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Mains power quality should
be that of a typical
commercial or hospital
environment. If you require
IEC 61000-4-11
0 % UT; 1 cycle
70% UT for 25/30
cycle
at 0°
0 % UT; 1 cycle
70% UT for 25/30
cycle
at 0°
continued operation during
power mains interruptions, it
is recommended that our
product be powered from an
0 % UT; 250/300
cycle
0 % UT; 250/300 cycle
uninterruptible power supply
or a battery.
(50/60 HZ)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
T
is the A.C. mains voltage prior to application of the test level.