ZS3 Service Manual Page 197 of 295
LED
#
LED
STATUS
Definition
1
AC Present
Solid: System has AC Mains applied and 5V
standby is active.
OFF: No AC detected, circuit breaker off.
2
Battery
OFF: No battery installed, or completely
depleted, or system is being used on battery.
Blinking: Battery charging (or reconditioning)
in-process.
Solid: Battery is fully charged.
3
“Controller
Status”
Slow Blinking: When power button is
pressed
Fast Blinking: Scan module is on and is
waiting for application software
Solid: System is on and ready for use.
4
“Interlocks”
Solid: Module docked and cover “on”.
5
Vwall
Solid: AC/DC 24VDC output is active.
OFF: No AC or circuit breaker tripped.
6
Vcart”
Solid: 12V for cart peripherals.
OFF: No power to cart assemblies.
7
“Vdock”
Solid: 12V for Scan Module.
OFF: No power to Scan Module.