Under “Setup | DICOM” verify the AE Title assigned to the ZS3 system is correct.
Select the Device List for the destination and hit “Ping”. If it fails, check IP address
and make sure DICOM server is setup and active on the network. If it still fails,
check with IT to confirm information entered is correct.
Under Setup | DICOM | DICOM Service | Storageverify DICOM AE Title and
port number are correct. Select the Service list and hit “Verify”. If it fails, check
with IT to confirm information entered is correct.
Under Setup | DICOM | DICOM Service | Print” verify AE Title and port number
are correct. Select the Service list and hit “Verify”. If it fails, check with IT to
confirm information entered is correct.
Using another device (PC, etc.) on same domain on hospital network, perform a
“PING” operation to the IP address assigned to the system in the Network
configuration page, to test for TCP/IP connectivity.