Check any short-circuits of the Y capacitor on power unit. Replace a new
one if any portion defective.
Check any broken of the Power Unit. Replace a new one if any portion
defective.
Inspect mains wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring
abnormalities. Notify the user or owner to correct any deviations. As a work
around, check the other outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console.
Inspect mains wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if situation
cannot be corrected, submit a Safety Failure Report to document the system
problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for
analysis and disposal.
1000 μA Single Fault Condition
IEC60601-1:: 500 μA Normal Condition
1000 μA Single Fault Condition