ZS3 Service Manual Page 291 of 295
ELECTRICAL SAFETY INSPECTION
8- Patient Auxiliary Current
overview
Patient Auxiliary currents are measured between any selected ECG jack
and the remaining selected ECG jacks. All measurements may have either a
true RMS or a DC-only response.
TEST PROCEDURE
Prepare
1) From the MAIN MENU, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
2) Attach the patient leads to the 601PRO ECG jacks.
3) Define the Lead Types from the View Settings Option.
4) Press shortcut key 8. The Patient Auxiliary Current test is displayed, and the test
begins immediately. Display values are continuously updated until another test is
selected.
5) Press SOFT KEYS 1-4 to select leakage tests
6) Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied
part leakage current:
7) Modify the configuration of the front panel outlet by pressing the appropriate
SOFT KEY on the 601PRO:
8) Press the print data key at any time to generate a printout of the latest
measurement.
Figure 4 patient Auxiliary Current