ZS3 Instructions for Use
119 | P a g e
its accessories or cause electric shock.
If the system is powered off improperly during operation, it may result in data damage of
the system hard disk or system failure.
Please use the ultrasound gel compliant with the relevant local regulations.
The cover contains natural rubber that can cause allergic reactions in some individuals.
In normal diagnostic ultrasound mode, there is no danger of a normal-temperature burn;
however, keeping the probe on the same region of the patient for a long time may cause
such a burn.
Do not use the system to examine a fetus for a long period of time.
When using intra-cavity probes, do not activate the probe outside the patient’s body.
Power off the system if you will not use the system for a long period of time (including
storage/ transportation condition), and you should not allow the system in standby status,
otherwise the batteries will be out of power and permanently damaged.
If you will not use the system for a long period of time, please periodically check the
battery and ensure that the battery power is sufficient. Otherwise, the battery may
permanently be damaged. If any fault is find during checking, please contact Mindray
Customer Service Department or sales representative.
To dispose of the system or any part, contact Mindray Customer Service Department or
sales representative. Mindray is not responsible for any system content or accessories
that have been discarded improperly.
DO NOT touch the Signal I/O ports if in contact with the patient; otherwise patient injury
may result.
Do not allow the patient to contact the live parts of the ultrasound system or other devices,
e.g. signal I / O ports. Electric shock may occur.
When faults occur during installing or leading the outside protective wire, the ultrasound
system should be powered by the internal power supply.
Do not remove any of the System covers other than the module cover. Other than the
scan module there are no user-serviceable parts internal to the system. Only trained
Mindray service personnel should access the system’s internal electronics.
The ultrasound system contains no operator-serviceable components within the
enclosures. To avoid electrical shock, do not remove covers. As with any other electrical
equipment, always observe care when operating this instrument. For service issues,
contact Mindray Technical Support. Failure to follow these restrictions may void your
warranty or service contract coverage.
To reduce the risk of electric shock, DO NOT connect the ZS3 system input or output
connections to equipment that is not properly connected to an Earth ground
To achieve proper grounding reliability, the ultrasound system power plug must be fully
inserted into a receptacle marked “hospital grade.” Do not remove the grounding wire. If
there is any question of power outlet or power cord integrity, do not proceed. Obtain
qualified assistance
To maintain proper grounding reliability, use only Mindray-recommended peripherals and
accessories. Use of non-specified peripherals and accessories could result in risk of
electrical shock or injury
The ultrasound system represents a potential explosion hazard if used in the presence of
flammable anesthetics.
The system does not contain a user-serviceable lithium ion battery.
The optional ZPAK cart battery is not a user serviceable item. Contact Mindray Technical