Configuration
AutoEF is a purchased license. Please confirm AutoEF has been configured before use.
Operations are as follows:
1. Press <Service> key to enter User Diagnostics Panel interface.
2. Click [License Info] to enter License Information interface, and check if AutoEF is
displayed on Features list.
Once configured, AutoEF is measured by entering Cardiac exam type.
Basic Procedures for AutoEF measurements
1. Press <New patient> key to start a new patient.
2. Select the proper probe and exam preset. Scan cardiac image.
Acquire an apical two-chamber view to measure the EDV(A2C) and ESV(A2C) and/or
Acquire an apical four-chamber view to measure the EDV (A4C) and ESV (A4C).
3. Press Freeze.
4. Press <Calc> key to enter measurement mode and select Auto EF or press the
programmable key assigned to Auto EF.
5. System will pause and then display traces on the ED and ES frames.
a) OLEDs are available to toggle between Systolic and Diastolic Frame by pressing the
OLED. Turn the dial to adjust the frame to be used for calculation.
b) Traces can be modified by pointing to the segment of trace to be adjusted. Caliper
points will become highlighted at which time the trackball can be used to drag the trace
to the desired position.
6. When data is acceptable, press ACCEPT OLED and data will be stored to the report.
7. If the height and weight have already been entered, the SV, EF, SI, EDV Index and ESV
Index are calculated. ECG obtains the HR automatically. If ECG is not connected, HR can
be selected on the menu.
Note: Auto EF can also be launched using a programmable key when the image is live. When
desired, press the Auto EF programmable key. The image will freeze and the above behavior will
follow.