20
To enter a value for x, y, u’, or v’
10.
Press [
/D-OUT] as necessary, so that the digit you want to set
next starts blinking (becomes active).
11
. Press [] or [] as necessary to change the value of the active
digit.
12.
Repeat steps 10 and 11 as necessary to set values for all digits.
When you have entered values for all digits, the entire
value (all digits) will blink.
You can cancel the entry procedure by pressing the
[CALL] key.
13
. Repeat steps 4 to 12 as necessary to set values for all param-
eters.
You can cancel the entry procedure by pressing the
[CALL] key.
14
. Recheck you entries to be sure that they are correct, and then
press [SET] to register them.
The CHROMA METER writes the entered values into
the target memory, so that they become the target val-
ues.
You can view the currently set target values by pressing [CALL].
If any of the values you have entered are outside of the permissible range for the corre-
sponding parameter, the CHROMA METER displays the “Err A” error message and then
cancels SET mode (without registering any of the entered values).