3 Setting Parameters
19
No. 99MBC109B
z
Sum/Difference calculation display (set value: 1)
Displays the Sum/Difference calculation for thickness or step measurement.
Tips
Set sum (A + B) or difference (A - B) by parameter number 14.
Connect Linear Gages that have the same resolution to INPUT A and INPUT B.
z
Dual display (set value: 2)
In Peak mode measurement, the peak value and the counter value of one
Linear Gage are displayed simultaneously. The peak value is displayed on
Display A, and the counter value is displayed on Display B.
You can switch between INPUT A and INPUT B using the BANK function.
Tips
For details about peak value switchover, see "4.5.1 Procedure for Switching Peak
Mode" (page 28).
For details about BANK switchover, see
"4.2 Switching the Displayed BANK (2-Axis
Models Only)" (page 25).
z
Memory display (set value: 3)
Saves the current value on Display A in Display B. In addition to the most
recently saved data, the maximum value and the minimum value from past data
can also be read out. The saved data will not cleared even if the power switch is
turned off.
Tips
For details about the procedures for saving, reading out, and clearing memory, see
"4.8 Registering, Reading Out, and Clearing the Memory" (page 34).
The memory unit is shared between BANK1 and BANK2. Connect Linear Gages that
have the same resolution.
The B_HOLD signal can be used to write to or clear the memory externally.
z
Speed display (set value: 4) (EH-102P/EH-102Z/EH-102S only)
Displays the moving speed of the Linear Gage's spindle simply.
In addition to the current speed, it displays the maximum speed when the
maximum value (MAX) is selected in Peak mode.
Tips
Specify the speed sampling cycle by parameter number 18.
The value is displayed in mm/sec. The display of the last 1 to 3 digits may be xed
depending on the speed sampling cycle.
The maximum speed in the reverse direction is displayed when the minimum value (MIN)
is selected in Peak mode.
This type is not applicable to feedback control.