4. SETTING UP THE MEASURING CONDITIONS
4.5.3.3 F2: Setting the GO/NG judgment criteria
Set the GO/NG judgment criteria according to the tolerance judgment method: (Lower limit
value + Upper limit value), (Multi-stage selection: 7 stages), and (Target value + tolerance),
whichever is specified in the basic setup. If “Using the abnormal value elimination function”
has been specified, the abnormal limit values should be set prior to setting the GO/NG
judgment criteria.
In this example assume that the machining target value is 12.5
+/- 0.01
mm, and that all the
abnormal limits (lower and upper) and GO/NG judgment criteria are canceled (set to 0).
1) Setting the abnormal limit values
Set as follows: Lower abnormal limit = 12.48 mm, Upper abnormal limit = 12.52 mm.
PROG
.4
12
8
PROG
Step 1: The previously set lower abnormal limit is
displayed.
Enter “12.48”.
PROG
Step 2: Press the
ENT
key.
The setup data will be saved in memory and
operation automatically proceeds to the setting for
the upper abnormal limit.
Step 3: Enter “12.52”, which is the setup data for the
upper abnormal limit.
.5
12
2
PROG
PROG
Case of
(Lower limit value and Upper limit value)
Step 4: Press the
ENT
key.
The setup data will be saved in memory
and operation will automatically proceed to
the GO/NG judgment criteria setting.
The method of GO/NG judgment varies
with the contents of the basic setup.
PROGPROG
Case of (Multi-stage selection) Case of (Target value + tolerance)