12
2 Basic Measuring Methods
No. 99MAF600B
2.1 Preparing Measurements
Perform "Origin and reference setup" and "Probe diameter setup" before measuring.
2.1.1 Setting the Origin and the Reference
Set the origin or the reference for measuring height. The length from this position will be measured as
the height. This product supports the ABS measurement system and the INC measurement system.
Use the method that matches your application.
y ABS measurement system (absolute value measurements)
The height relative to the set origin is directly displayed.
For details about setting the origin, see
"■Settingtheorigin(ABSmeasurementsystem)"on
page 13.
Example: The surface plate is set as the origin at 0 mm. Example: Surface s1 is set as the origin at 50 mm.
s1
50 mm
0 mm
y INC measurement system (comparative measurements)
The height relative to the set reference (0 mm) is displayed.
For details about setting the reference, see
"■Settingthereference(INCmeasurementsys-
tem)" on page 16.
Example: Surface s1 is set as the reference (0 mm).
s1 (0 mm)
Tips
When the power supply to the product is terminated by removing the batteries or unplugging the AC adapt-
er (optional accessory), the set origin or reference will be erased. In this case, it must be set again.
If the workpiece is large, and it is necessary to move the main unit, grab the base with your right hand and
the carrier grip with your left hand to move it.