41
3SpecicUsageApplications
No. 99MAF600B
3.2.1 Setting the Upper and Lower Tolerance Limits
As an example, the procedure for setting the upper tolerance limit [+100.010mm] and lower tolerance
limit [+99.995mm] is explained below.
Tips
• Thesettolerancelimitsareretainedinmemoryevenifthepoweristurnedo.
• If the upper tolerance limit is set to a value that is lower than the value set for the lower tolerance limit,
there will be an error.
1
Press [MODE].
» Characters and icons will be displayed in the bottom row
of the LCD.
2
Press [Inside diameter measurement] to select [TOL.] (tolerance setting).
M1 M2
» [U] and the current upper tolerance limit will be displayed
on the LCD.
3
Press [→] to make [+] blink.
» [+] will blink.
If[-]isblinking,press[↑]or[↓]tochangethedisplayto[+].
4
Repeatedly press [→]untilthevalueinthehundredsplaceashes.
» Thevalueinthehundredsplacewillash.