15
2 Basic Measuring Methods
No. 99MAF600B
5
Display [5] in the ones place using the same procedure described in steps
3
and
4
.
6
Set the origin.
1 Loosen the clamp knob.
2 Loosen the clamp screw.
3 Turn the probe unit up/down wheel, and slowly bring the stylus into contact with the top surface of the gauge
block until there is a "beep" sound.
1
2
3
3
» The origin will be set, and
the height relative to the set
origin will be displayed in the
top row of the display.
Tips
For measurements using a lever-type dial gauge (optional accessory) or dial
gauge (optional accessory), bring the stylus into contact with the gauge block
without loosening the clamp knob, and after aligning the hand to the zero point,
press [ENTER]tonishthesetup.
7
Conrmthesetorigin.
1 Slowly move the stylus away from the top surface of the gauge block, and bring them into contact again by
turning the probe unit up/down wheel until a "beep" sound is made.
2 Check that [25.000] is displayed on the LCD (the origin setup is complete).
Tips
Iftheoriginissignicantlydi󰀨erentfrom[25.000],performsteps
1
through
7
again.