5 parameters of the unit
Note: All technical data are related to 22 °C ambient temperature and ± 10 % voltage swing (if not specied otherwise).
For other parameters see section 5.1 – Electric connections
*) Approx. 50 % of the tray area can be lled in a way a uniform air circulation is enabled inside the chamber.
**) Standard type is up to 250 °C, optional type is up to 300 °C
***) Standard type is up to 70 °C, optional type is up to 99,9 °C
****) Supply voltage is indicated on the type lable of the unit
*****) Temperature deviations were measured in accordance with DIN 12 880
+) Two-door pass-through model; applies to Venticell only
ECOCELL(EC), DUROCELL(DC), VENTICELL(VC),INCUCELL/INCUCELL V (IC/ICV)
Technical data
Interior of stainless steel
material DIN No 1.4301
volume EC, VC, IC / ICV
DC
cca liters 22
22
55
55
111
111
222
222
222/2+)
404
-
404/2+)
707
-
707/2+)
width mm 240 400 540 540
540 540 540 940 940
depth EC, DC, VC
IC / ICV
mm 320 390
370
390
370
540
520
540 540
520
540 540
520
540
-
height mm 295 350 530 760
760 1410 1410 1410 1410
Tray bearing plate, width x depth
racks
distance of the wiring
standard equipment
mm
No
mm
pcs. included
185x265
4
60
2
380x335
4
70
2
520x335
7
70
2
520x485
10
70
2
520x485
10
70
2
520x485
19
70
2
520x485
19
70
2
920x485
19
70
2
920x485
19
70
2
Maximal permissible load to the
trays *)
per tray
total load
kg/tray
kg/unit
10
25
20
50
20
50
30
70
30
70
30
100
30
100
50
130
50
130
Door psc. 1 1 1 1
2 1 2 2 4
External dimensions
(including door and handle)
width
depth
height (incl.Feet F and Rolls R)
Ø of the air branch outer / inner
mm
mm
mm
mm
406
580
604F
52/49
620
640
680F
52/49
760
640
860F
52/49
760
790
1090F
52/49
760
790
1110F
52/49
760
790
1910R
52/49
760
790
1910R
52/49
1160
790
1910R
52/49
1160
790
1910R
52/49
Package dimensions (three
layers carton)
width
depth
height (incl. palette)
mm
mm
mm
465
665
655
700
730
880
830
730
1050
830
860
1280
830
860
1280
830
860
2070
830
860
2070
1230
860
2080
1230
860
2080
Weight
nett
brutt
kg
kg
31
36
55
66
75
87
100
116
105
121
150
175
150
185
215
240
230
255
Elektricity
power W input [stand by] 5 5 5 5
5 5 5 5 5
- mains 50/60 Hz max. power input kW
VC
DC
IC / ICV
EC
0,96
0,92
0,24/0,96
0,92
1,3
1,2
0,3 / 0,7
1,2
1,9
1,8
0,3 / 0,7
1,8
1,9
1,8
0,5 / 0,7
1,8
3,7
3,7
-
0,9 / 1,3
3,6
5,5
4,9
-
0,9 / 1,3
4,5
7,2
voltage V ****) EC, DC, VC
IC / ICV
230
230
230
230
230
230
230
230
400/3NPE
400/3NPE
230
400/3NPE
400/3NPE
230
400/3NPE
-
current A VC
DC
IC/ICV
EC
4,2
4,0
1,1 / 4,2
4,0
5,6
5,2
1,3 / 3
5,2
8,3
8,3
1,3 / 3
7,8
8,3
7,8
2 / 3
7,8
5,2
5,2
-
3,9 / 5,6
7,8
7,9
7,8
-
3,9 / 5,6
7,8
10,4
voltage V ****) EC, DC, VC
IC / ICV
115
115
115
115
115
115
115
115
115/3PE 115/3PE
115
115/3PE 115/3PE
230
115/3PE
-
Temperature data
Working temperature
(beginning of the regulation)
od 10 °C above ambient temp. to °C
od 5/10 °C
above ambient temp. to °C
od 5 °C above ambient temp. to °C
VC **)
IC / ICV ***)
EC
DC
250 /300
70 / 99,9
250
125
250 / 300
70 / 99,9
250
125
250 / 300
70 / 99,9
250
125
250 / 300
70 / 99,9
250
125
250 / 300 250 / 300
70 / 99,9
250
-
250 / 300 250 / 300
70 / 99,9
50
-
250 / 300
-
-
-
The temperature
deviations*****) according to
DIN 12 880, Section 2, from the
average value of the working
temperature in the space of the
chamber, when the ventilation
ap and doors are closed
VC (>50 °C) space deviation
time deviation
cca (±) % Average value
In the space of the
chamber cca (±) ºC
1,1
0,3
1
0,4
1
0,4
1
0,4
1
0,4
1,8
0,9
1,5
0,4
2,5
0,4
2,5
0,74
IC/ICV
at 37 °C
space deviation
time deviation
cca (±) °C
do (±) °C
1,1/0,4
0,4/0,1
<0,5/≤0,3
≤0,2
<0,5/≤0,3
≤0,2
<1/≤0,3
≤0,2
<1/≤0,8
≤0,2
<1,5/≤1,5
≤0,2
-
-
EC space deviation
time deviation
cca (±) % Average value
In the space of the
chamber cca (±) ºC
2,7
1,0
2
≤0,3
2
≤0,8
2
≤0,8
2,5
≤1
3,5
≤1
-
-
DC
at 100°C
space deviation
time deviation
cca (±) % Average value
In the space of the
chamber cca (±) ºC
3,1
0,4
2
≤0,3
2
≤0,8
2
≤0,8
-
-
-
-
-
-
Temperature
rise time with
closed air ap
and at the voltage
of 230 V ± OV
250 °C VC
EC
cca min.
28
54
49
59
53
60
70
99
33 58
85
43 64
95
50
-
100 °C DC 34 41 48 50
- - -
37 °C IC/ICV 70/8 49 / 41 57 / 51 79 / 66
41 / 38 59 / 51 -
Heat emision at 250 °C EC, VC cca W 300 / 420 590 760 990
990 1940 1940 2550 2550
100 °C DC 140 380 490 630
- - -
37°C IC/ICV 20/70 30 45 45
65 85 -
Air exchange
rate at
150 °C VC
EC
cca/h 45
6
45
8
49
12
24
5
24 18
4
18 12
3
12
-
100 °C DC 6 8 12 5
- - -
37 °C IC/ICV 4 / 45 5 / 45 5 / 49 5 / 24
5 / 18 5 / 12 -

Table of Contents