PROGRAMMING A TPN THERAPY USING MANUAL PROGRAM MODE
NOTE: When operating the pump on battery power, the LCD screen dims after a period of time to
conserve battery power. Any keypress will illuminate the screen.
NOTE: Use the UP/DOWN ARROW keys to move highlighter. Use the numeric keys for data entry.
1. Turn the pump ON and wait for the self-test to complete.
2. Highlight PROGRAM from the SELECT menu and press YES.
NOTE: The following screen (“NEW PROGRAM”) will not appear if the previous RX has been cleared from
the pump. If this screen does not appear, skip to step #4.
3. Highlight NEW PROGRAM and press YES. Enter access code, if prompted.
4. Highlight TPN from the therapy selection menu and press YES. If any on-screen alerts appear, press
the YES key.
5. With Unidentied highlighted, press YES (this indicates Manual Programming Mode).
6. Press YES to leave the Delayed Start O, or press NO to set a delay and refer to the Delayed Start
section.
7. At Next? Press YES to continue.
NOTE: Bag Volume should include overll volume if pump priming and/or a KVO is programmed.
8. Enter Bag Volume, press YES.
NOTE: To force an auto-calculation, enter any 3 of the 4 entries: Vol TBI, INF RATE, UP/DN Ramp or TOT
TIME, the unknown eld will auto-calculate. Bypass the unknown eld by using the DOWN
ARROW key (typically the INF RATE). Enter the known elds pressing YES to conrm each. The
unknown eld will auto-calculate. Press YES to conrm this eld.
9. Enter Vol TBI (if known), press YES.
10. Enter INF RATE (if known), press YES.
11. Enter UP Ramp and DN Ramp (if any), press YES after each entry.
12. Enter TOT TIME (if known), press YES. Continue pressing YES if needed until the KVO eld is
highlighted.
NOTE: If a KVO rate of 0.1 to 10 mL is entered, the pump will give audible alerts every 20 minutes at the
end of the infusion. Whereas if zero KVO is entered the pump will alarm every 3 seconds at the
end of the infusion.
13. Enter KVO rate (if any), press YES.
14. Press YES at DONE?
TPN THERAPY