PRIMING THE SET
CAUTION: Disconnect set from patient’s access site before priming.
NOTE: Review program settings prior to priming.
NOTE: Filters must be primed in the upright position (do not invert). Refer to Instructions for Use in set
package.
BY GRAVITY:
NOTE: To remove air from IV bag, turn bag upside down after attaching the set and gently squeeze bag
while priming until air is eliminated from the bag.
1. Twist and remove breakaway tab from YELLOW Flow-Stop prior to connecting to solution container.
2. Squeeze YELLOW Flow-Stop between thumb and forenger to allow uid to ow through tubing.
BY PUMP:
NOTE: Therapy must be programmed and administration set must be installed into the pump prior to
pump priming. Sets with in-line ASV must be primed on the pump.
1. Release any clamps on the administration set.
2. Press the PRIME key from the “RUN TO START” screen.
3. Press and hold the PRIME key until set is free of air; release PRIME key to stop (each prime cycle
allows up to 6 mL of uid).
4. Press the YES key to exit priming.
CHANGING THE SOLUTION CONTAINER
1. Press the PAUSE key.
2. Change the solution container.
3. Highlight REPEAT RX, then press YES.
4. Highlight correct therapy if requested, then press YES.
5. If any on-screen alerts appear, press the YES key.
6. Note any on-screen ADVISORY and press YES to continue.
7.The program will auto-scroll (or conrm each eld by pressing YES).
8. Press RUN to start the infusion.